Aplisens Smart Differential Pressure Transmitter PRE-28 Smart

Aplisens

Write a Review

Configured Per Application – Call for Quote

SKU:
PRE-28 Smart

Product Overview

Pricing Available Upon Request
The Aplisens PRE-28.SMART is configured per application (DP range, process connection style, casing/IP rating, and approvals like Ex ia / Ex d / marine). For correct selection and accurate pricing, please call 865-900-8844 or email KSmith@CPSCalibrations.com.

The Aplisens PRE-28.SMART is a smart differential pressure transmitter with 4–20 mA output + HART for configuration, diagnostics, and integration into PLC/SCADA systems. It supports intrinsic safety (ATEX/IECEx) and explosion-proof (ATEX/IECEx) versions, with static pressure ratings up to 413 bar (configuration dependent).

  • Signal: 4–20 mA + HART
  • Accuracy: ±0.1% of calibrated range
  • Static pressure limit: up to 413 bar (C-type, option dependent)
  • Wetted parts: SS316L (diaphragm options include Au)
  • Approvals: Ex ia / Ex d (ATEX, IECEx), marine certificates (DNV, BV)
  • Multiple housings: IP65 / IP66 / IP67 / IP68 options (depending on casing)
  • Process connections: manifold mount (C / CH) or threaded (GP / PN), plus diaphragm seal option

Tell us your DP range, static pressure requirement, process connection style, and required approvals (Ex ia / Ex d / marine), and we’ll confirm the correct build for your application.

Feature Details
Model Aplisens PRE-28.SMART
Type Smart Differential Pressure Transmitter
Output / Communication 4–20 mA + HART
Power Supply 7.5…55 VDC (Ex ia: 7.5…30 VDC)
Resistance Required for HART Minimum 240 Ω
Accuracy ±0.1% of calibrated range
Long-term Stability ≤ accuracy for 3 years (for the basic range)
Thermal Compensation Range -25…80°C
Ambient Temperature -25…85°C (Ex ia version: -25…80°C)
Medium Temperature -25…120°C (over 120°C typically requires impulse line or diaphragm seals; up to 100°C for 413 bar static pressure option)
Electronic Damping 0…30 s
Wetted Parts SS316L
Diaphragms SS316L (Au option available)
Casing Material SS304 (optional SS316 depending on casing)

Measuring ranges (nominal FSO) & minimum set range:

No. Nominal Measuring Range (FSO) Minimum Set Range Rangeability Overpressure / Static Pressure Limit
1 0…70 bar 7 bar 10:1 C-type: 250 / 320 / 413 bar (options)
GP-type: 40 bar (for range 1)
2 0…16 bar 1.6 bar 10:1 C-type: 250 / 320 / 413 bar (options)
3 0…2.5 bar 0.2 bar 12.5:1 C-type: 250 / 320 / 413 bar (options)
4 0…1 bar 50 mbar 20:1 C-type: 250 / 320 / 413 bar (options)
5 0…0.25 bar 10 mbar 25:1 C-type: 250 / 320 / 413 bar (options)
6 -0.5…0.5 bar 0.1 bar 10:1 C-type: 250 / 320 / 413 bar (options)
7 -100…100 mbar 10 mbar 20:1 C-type: 250 / 320 / 413 bar (options)
8 -5…70 mbar 4 mbar 18:1 C-type: 250 / 320 / 413 bar (options)

Note: Overpressure/static pressure capability depends on process connection and selected options. For hot media, use an impulse line or diaphragm seal.

PRE-28.SMART pricing depends on configuration. To quote correctly, we typically confirm:

What We Need Examples
Approvals / Version /Exia (intrinsic safety) or /Exd (explosion-proof); /MR (marine DNV/BV)
Static Pressure Option /320 bar or /413 bar (C process connection only)
Nominal Range + Calibrated Range Select nominal range (FSO), then set the calibrated 4 mA to 20 mA range to match your process
Casing / Electrical Connection /PD (IP65 DIN), /PZ or /PZ316 (IP66 gland), /PM12 or /PM12S (IP67), /SG or /SGM (IP68 cable), /FL flying leads (Exd only), /PZ2 (Exd only)
Process Connection /C or /CH (manifold mount), /GP (G1/2" M), /PN (1/4" NPT F), diaphragm seal on Hi side (Lo side 1/4" NPT F)
Wetted Materials / Diaphragm SS316L standard; Au diaphragm option for C-type (as specified)
Gasket (C/CH) FPM Viton standard; /NBR or /PTFE (PTFE max static pressure 250 bar)
Extras /Tlen oxygen service; /NACE for C process connection; mounting brackets and impulse line adapters as needed

Call 865-900-8844 or email KSmith@CPSCalibrations.com with your DP range, static pressure, connection style, casing preference, and approvals and we’ll confirm the correct PRE-28.SMART configuration.