Product Overview
Description
Introduction
The SENSIT® GOLD G2 multi-gas detector is a highly advanced instrument designed to locate gas leaks and use in confined spaces. It is suitable for detecting combustible gases in oxygen-free environments, making it ideal for inert gas operations. The instrument can be configured for combustible gas readings in PPM, LEL and/or percent volume, and various toxic gases and oxygen.
Displays Up to 4 Gases Simultaneously
The SENSIT® GOLD G2 is extremely accurate, durable, and easy to use and maintain. It can display up to four gases simultaneously, including:
- LEL (% volume and PPM optional)
- Oxygen, Carbon Monoxide
- Hydrogen Sulfide
- Hydrogen Cyanide (optional)
It also has selectable gas types, such as methane (natural gas) and propane.
Partial List of Gases Sensed
The instrument can sense a partial list of gases, such as:
- Butane
- Carbon Monoxide
- Gasoline
- Hydrogen Sulfide
- Hydrogen Cyanide
- Methane
- Natural Gas
- Oxygen
- Propane
Standard Features
It has several standard features:
- Intrinsically Safe Design (certified)
- Internal Pump for Fast Sampling
- Audible “Tick” Control for Fast Leak Location
- Lowest Cost
- Long Life Sensors
- LED Warning Lights
- Bright Graphic Display
- Loud Audible Alarm
- Battery Strength Indicator
- Internal Memory
- Exceptional Water Protection
Applications
The SENSIT® GOLD G2 finds its applications in detecting combustible gas leaks, gas leak surveys, underground leak pinpointing, carbon monoxide investigations, confined space monitoring, combustible gas measurements, pipeline purging, method 21 emissions testing, and inert gas operations.
Certifications
It is certified by ATEX and UL.
Product Specification
Specification | Details |
---|---|
Size | 11.5” x 3” x 2.32” (29.2 x 7.6 x 5.9 cm) |
Weight | 1.2 lb. (544 g) |
Operating Temperature | -4 to 104° F (-20 to 40° C) |
Storage Temperature | 25° C to +65° C (-13° F to +149° F) |
Battery Life | Alkaline: Approximately 18 hrs. Continuous |
Standard Kit |
|
Additional Features
Feature | Details |
---|---|
Barhole Test Features | Peak Readings, Sustained Readings, Timed Test |
Adjustable Calibration Reminders | Calibration Due, Calibration Over-Due (Shows at Start-up) |
Gas Identification Alert Options | “NSR” - Non-Standard Response, “NSC” - Combustible Other Than Gas Calibrated to (4-5% Volume) |
Optional Features | Data logging Options: 100 Instant Event Saves, 1600 Auto Log Saves, 100 Bar Hole Tests |
Sensor Specifications
Type | Resolution | Range | Accuracy |
---|---|---|---|
PPM* | 1PPM or 10PPM | 0-10,000PPM | ±10% |
LEL | 0.1% up to 2% | 0-100% LEL | ±10% |
% NAT GAS | 0.1% | 5-100% GAS | ±5% |
% Propane | 0.1% | 2.2-100% GAS | ±5% |
O2 | 0.1% | 0-25% | ±0.2% or 10%** |
CO | 1PPM | 0-2000PPM | ±5PPM or 5%** |
H2S | 1PPM | 0-100PPM | ±2PPM or 5%** |
HCN | 1PPM | 0-30PPM | ±1PPM or 5%** |
*PPM Optional
**Whichever is greater
Product Videos
Videos Hide Videos Show Videos
-
SENSIT GOLD G2 Multi-Gas Detector Introduction
An introductory overview of the SENSIT® GOLD G2 Combustible Ga...
-
SENSIT GOLD G2 Multi-Gas Detector Quick Start Guide
SENSIT® GOLD G2 Gas Leak Detector Quick Start. This video help...
-
SENSIT GOLD G2 Calibration
How to calibrate the SENSIT GOLD G2 Multi-Gas Detector SENSIT...
-
SENSIT GOLD G2 Calibration Using The SMART-CAL Calibration Station
How to calibrate the SENSIT® GOLD G2 Multi-Gas Detector using ...
-
SENSIT GOLD G2 Tick Feature
SENSIT GOLD G2 Tick Feature for gas leak search and pinpointing.
-
SENSIT GOLD G2 Leak Search Mode
SENSIT GOLD G2 Leak Search Mode
-
SENSIT GOLD G2 Purge Mode
SENSIT GOLD G2 Purge Mode for gas pipe systems.
-
SENSIT GOLD G2 Bar Hole Test Mode
SENSIT GOLD G2 Bar Hole Test Mode.