Description
The NAGCO N012-3022-010 Benzene Separation Tubes are designed for measuring benzene (C6H6) concentrations using a 9.8 eV standard lamp. These separation tubes provide a broad typical measurement range of 0.1–1000 ppm for benzene measurement applications.
Measurement time and required sample volume vary based on temperature. NAGCO specifies measurement times from 40 to 150 seconds and sample volumes from 300 to 1125 mL depending on the temperature at which the measurement is performed.
Key Features
- Designed for benzene (C6H6) measurement
- Typical measurement range: 0.1–1000 ppm
- Designed for use with a 9.8 eV standard lamp
- Temperature-dependent measurement time and sample volume
- No effect on reading from 0–95% RH under specified conditions
- Yellow → Brown → Green color change
- 3-year unopened storage life under specified storage conditions
Specifications
- Manufacturer: NAGCO
- Part Number: N012-3022-010
- Target Gas: Benzene (C6H6)
- Typical Range: 0.1–1000 ppm
- Standard Lamp: 9.8 eV
- Temperature Range: 2–40°C (36–104°F)
- Humidity: No effect on reading from 0–95% RH
- Color Change: Yellow → Brown → Green
- Storage Life: 3 years unopened in darkness at 5–25°C (40–77°F); refrigeration preferred
- Opened Tube: Use within 1 hour to avoid loss of capacity
Measurement Time & Sample Volume
- 2–10°C (36–50°F): 150 seconds / 1125 mL
- 10–15°C (50–60°F): 90 seconds / 675 mL
- 15–30°C (60–86°F): 60 seconds / 450 mL
- 30–40°C (86–104°F): 40 seconds / 300 mL
Calibration & Use
Calibration should be performed at the same temperature as the measurement. NAGCO recommends recalibration when changing tube batches. Humid, clean air drawn through the tube before measurement can reduce VOC capacity.
Cross-Sensitivity & Interferences
The N012-3022-010 has documented cross-sensitivity to several compounds. NAGCO identifies responses from compounds including chlorobenzene, trichloroethylene and perchloroethylene. Other compounds may affect tube response or capacity. Refer to the NAGCO datasheet for complete cross-sensitivity information before use.
Important: Drawing humid air through the tube for extended periods or drawing liquid water through the tube may damage the instrument. Each tube contains 3 mg of chromium compounds. Always follow the NAGCO datasheet and instrument operating instructions when performing measurements.
View NAGCO N012-3022-010 Benzene Separation Tube Datasheet (PDF)